Product ID: DP-25 25MHz Differential Probe 3 RANGE ATTENUATOR. 2 POWER SOURCE. 1000V INPUT CAT III LOW BATT. AUTO OFF. SEPARATING DESIGN.CONVENIENT, DURABLE. MEET CE, IEC 1010, EN 61010, UL 3111. Band width (-3dB, 50Ω load): DC-25MHz . ( x 20: DC~15MHz) Accuracy : +/- 2% at 20-30'C 70% RH after 20 minutes warm up. Attenuation : x 20, x 50, x 200 ( Into 1M Ohm scope ) ; x 40, x 10 , x 400 ( Into 50 Ohm load ) Maximum operation Voltage ( DC + peak AC ) : <= 140V at x 20 <= 350V at x 50 <= 1300V at x 200 Maximum input Differential Voltage : 1300V ( DC + peak AC ) or 1000V RMS Maximum input Voltage to Ground : 1000V ( DC + peak AC ) or 600V RMS Common Mode Rejection Ratio ( CMRR ): 60Hz : >10,000 : 1 100Hz: >1,000 : 1 1MHz: > 300 : 1 Noise :<= 2 mV rms ( Into 50 Ohm load ) Input Impedance : 2M Ohm, 2.3 PF between inputs and ground. 4M Ohm, 1.2 PF between inputs. Power :One internal 9V alkaline battery or external 6V-9V DC.
Specifications:
FEATURES
FEATURES
The DP-25 FET input differential probe provieds a safe means of measuring circuits with floating potentials up to 1000 V ( DC+ peak AC ) from ground and 1300V ( DC + peak AC ) differential.
The DP-25 converts the high voltage differencial input signal to a low voltage ground refereced signal for display on any Oscilloscope.
The output BNC of DP-25 is calibrted to drive a high impedance ( 1M Ohm) load.
INSTRUCTION FOR USE
Connect the output BNC of DP-25 to the input BNC of the Oscilloscope by the accessory BNC cable.
Adjust the vertical offset of the Oscilloscope if necessary.
Set the select proper range of the DP-25 and the V / DIV of the Oscilloscope according to the scale conversion chart.
Scale conversion chart : The effective V / DIV is the attenuation factor of x 20, x 50, x 200 multiplied by the scale factor of the Oscilloscope. It will be twice when the 50 Ω load was used. For example, with the range set at x 200,
and the scope set to 0.5 V / DIV, the effective V / DIV equals 200 x 0.5 or 100 V, with the 50 Ohm load was used, it becomes 200V, the power consumption will increase too.
NOTE:If the voltage of the input signal exceeeds the linear range of the setting range.The signal output of
the DP-25 would not accurately, the wave form display will be cut off.
WARNING
WARNING
Do not use DP-25 above 1000v ( DC + peak AC ) between ground and the input or 1,300V ( DC + peak AC ) between the input lead.
Do not operate DP-25 in wet or damp condition.
Do not operate DP-25 in an explosive atmosphere.
Do not immerse DP-25 in liquids.
Do not operated DP-25 without covers.
Please change the battery when the "LOW BATT " LED is lighted. At this time DP-25 can operate but not guaranteed the accuracy.
DP-25 can not operate if both POWER and LOW BATT LED are not light.
FEATURES
The DP-25 FET input differential probe provieds a safe means of measuring circuits with floating potentials up to 1000 V ( DC+ peak AC ) from ground and 1300V ( DC + peak AC ) differential.
The DP-25 converts the high voltage differencial input signal to a low voltage ground refereced signal for display on any Oscilloscope.
The output BNC of DP-25 is calibrted to drive a high impedance ( 1M Ohm) load.
INSTRUCTION FOR USE
Connect the output BNC of DP-25 to the input BNC of the Oscilloscope by the accessory BNC cable.
Adjust the vertical offset of the Oscilloscope if necessary.
Set the select proper range of the DP-25 and the V / DIV of the Oscilloscope according to the scale conversion chart.
Scale conversion chart : The effective V / DIV is the attenuation factor of x 20, x 50, x 200 multiplied by the scale factor of the Oscilloscope. It will be twice when the 50 Ω load was used. For example, with the range set at x 200,
and the scope set to 0.5 V / DIV, the effective V / DIV equals 200 x 0.5 or 100 V, with the 50 Ohm load was used, it becomes 200V, the power consumption will increase too.
NOTE:If the voltage of the input signal exceeeds the linear range of the setting range.The signal output of
the DP-25 would not accurately, the wave form display will be cut off.
Certificate:
CE
Contact Detail